Operation instructions
• Instrument should be plugged on the power
according to the requested technical parameters and
connected to the earth leakage circuit breaker.
• In order to calibrate the screen, switch on the system
and follow the instructions below:
a) Tap in any part of the screen except for the icons
until you will hear a beep.
b) Press the small crosses that will appear on the
screen.
c) Once the screen has been calibrated, get back to the
main menu.
BEFORE THE TREATMENT:
1) Select the programme according to the area the
patient wants to treat. Meanwhile, we recommend
recording the current status of the patient and the
proposed programme. This is to identify the difference
before and after treatment and as guidance for next
treatment.
2) Soften the surface skin to be treated with a steamer
or hot pad for 10 min. And then massage the muscle
for 5-10 min. with a scrub cream or oil. Meanwhile, it is
recommended to create a relaxing and comfortable
atmosphere.
3) Wash the excess cream or oil after the massage.
Rub the skin surface with a damp sponge.
4) Apply some nutritive lotion or gel on the electrode
surface. As shown in the diagrams, use elastic belts
for a better xation of the electrode on the treated
skin.
5) Check that the electrodes are correctly placed. Also
make sure that elastic belts are properly tightened and
that the electrodes are in contact with the skin.
DURING THE TREATMENT:
1. Switch on the unit and select the language rst.
Then select the desired treatment.
2. Once the treatment is chosen, an interface will show
you the details of the corresponding programme and
where the electrodes should be applied. Once the
electrodes are well-placed, press "OK" for next step.
3. Select PLAY to start working and time will begin to
count down at the same time.
4. The corresponding channels should also be
switched on. To start working, intensity should be
adjusted at the minimum position. Later the intensity
can be adjusted according to the reaction of the
muscle.
5. Once the programme is nished, please remove the
electrodes and clean the treated skin area with a wet
cloth or tissue.
6. Clean the unit and accessories after each treatment.
The accessories should be sterilised as well.
The muscle strengthening programme allows selecting
the different parameters for each programme:
- Interval: once the default time is nished, the unit
will have a pause of one minute. If you adjust the
interval parameter, you can modify the pause time.
- Group: Once the working and the pause time have
nished, a cycle is completed. By adjusting the group
parameter, you can select the number of cycles you
desire.
Notes for the placement of the electrodes:
• One group of electrode consists of 2 pieces of
electrodes with the same number and reverse polarity.
• Place the electrodes according to the polarity (+ or -).
Otherwise the treatment effect will get reduced and
the patient will feel uncomfortable.
• Each pair of electrodes should be symmetrically
placed according to the schemes shown; the distance
between both electrodes should not be less than 2 cm.
entre ambos electrodos no debe ser menor de 2 cm.