SimplyGo Mini User Manual
Visual, Audio
Indicators
1 beep
that repeats every
16 seconds
36
Description
Low Battery Alarm
This alarm occurs when
approximately 10 minutes
of battery life remains�
Remaining battery life is
dependent on your device
settings and your activity
level�
Press the yellow Alarm
silence button on the
screen to temporarily
silence the alarm, or press
the Home button on the
device to return to the
Home screen, and the 2nd
screen will appear with
the Low Battery symbol
flashing on the screen�
Note: This alarm could
take up to 46 seconds to
activate from the time
the alarm condition is
present�
Warm Up Indicator
This symbol indicates to
the user that the O
has not yet been sampled
because the system needs
to stabilize� Its duration is at
least 14 minutes�
What To Do
Replace the battery
or connect to a
power source�
This symbol
appears onscreen
after initial startup
and clears from the
screen once the
device has reached
purity
2
the minimum rated
Oxygen Purity�
Contact your
equipment provider
if it persists on
screen for several
hours after startup�