Setting The Mixing Time; Changing The Basic Settings; Operation; Insertion Of The Capsule - 3M CapMix ESPE AG Instrucciones De Servicio

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 18
SEITE 3 - 148 x 210 mm - 063/0/-/51+ F -

5.1 Setting the mixing time

Mixing times in the range of 1sec to 39 sec may be
selected. By pressing the keys »+« or »-« the device
increases or decreases, respectively, the mixing
time by 1sec. Depressing and holding the »+« or »-«
keys, the device continues to count up or down,
respectively.

5.2 Changing the basic settings

Set the mixing time as described above, then depress
one of the program buttons, »P1« - »P4«, until two
short signals are issued.This confirms that the
desired settings have been stored in the selected
program.

6. Operation

6.1 Insertion of the capsule

Open and fold back the cover of the CapMix device.
Press down on the release button to release the capsule
holder.
Keep the release button depressed and push to the left
to open the capsule holder.
Place the activated capsule between the two cups of
the holding mechanism and release the release button.
The device clamps the capsule automatically. Insert
capsules with tips such that the tip points towards the
left side.
Do not operate the device without a capsule inserted!
SCHWARZ
- 03-0755 (sr)

6.2 Mixing

Close the protective cover and press the »Start/Stop«-
button.This starts-up the mixing process.The countdown
of the mixing time is shown on the time display.
After the mixing time has elapsed, the latest time
setting entered is re-displayed so that there is no need
to re-enter any value if the same time is to be used.
The selected program can be interrupted at any time
by pressing the »Start/Stop«-button, upon which the
program returns to its original setting.
CAUTION: For safety reasons, the unit cannot be
started up unless the cover is closed. Attempting to
operate the unit with the cover being open causes
the time display to flash and an error message to be
issued. Opening of the protective cover during mixing
is associated with a high risk of injury. Hence, the
mixing process is aborted.This may render the capsule
useless! Therefore, never open the cover before the
mixing process is completed and the capsule holder
has come to a complete stop!

6.3 Removal of the capsule

Open the lid of the CapMix unit.
Open the capsule holder by pressing the release
button down and towards the left, and remove the
capsule.

7. Notes

Capsule materials of third-party manufacturers (including
amalgam capsules) may be mixed with the CapMix
device provided they are suitable for the device in terms
of weight and dimensions (please refer to the section,
Technical Data).The optimal mixing time for mixing with
the CapMix device should be inquired from the capsule
manufacturer.
Pay attention to capsule leak tightness when mixing.
Capsules that are not leak tight contaminate the device
with mercury.

8. Disposal

This unit is made from materials which can be recycled
and reused. Special recycling companies possess
the technology to disassemble old units into individual
components and reuse recyclable components leading
to a substantial reduction in the amount of materials to
be disposed. Please inquire locally about the available
recycling resources.
3

Publicidad

Capítulos

Tabla de contenido
loading

Tabla de contenido