Beurer BC 40 Manual De Instrucciones página 103

Ocultar thumbs Ver también para BC 40:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 39
Table 3
Guidance and manufacturer´s declaration – electromagnetic immunity
The BC40 Wrist Digital Blood Pressure Monitor is intended for use in the electromagnetic environment specified below. The customer or the user of
the BC40 Wrist Digital Blood Pressure Monitor should assure that it is used in such an environment.
Immunity test
IEC 60601 test
level
Conducted RF
3 V
150 kHz to
rms
IEC 61000-4-6
80 MHz
Radiated RF
3 V/m 80 MHz to
IEC 61000-4-3
2,5 GHz
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic is affected by absorption and reflection from structures,
objects and people.
Compliance
Electromagnetic environment – guidance
level
Portable and mobile RF communications equipment should be used no closer to any part
of the BC40 Wrist Digital Blood Pressure Monitor, including cables, than the recommen-
ded separation distance calculated from the equation applicable to the frequency of the
transmitter.
Recommended separation distance:
3
5 .
d
[
]
=
V
1
Not applicable
3
5 .
d
[
]
=
V
1
7
d
[
]
=
3 V/m
E
1
where p is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer and d is the recommended separation distance in metres
(m).b
Field strengths from fixed RF transmitters, as determined by an electromagnetic site
survey,a should be less than the compliance level in each frequency range.b
Interference may occur in the vicinity of equipment marked with the following symbol:
P
P
80 MHz to 800 MHz
800 MHz to 2,5 GHz
P
103

Publicidad

Tabla de contenido
loading

Tabla de contenido