Section 3 - Operator Control Panel
®
3.0 Operator Control Panel
The operator controls are identified in figure 4.
3.1 POWER Switch
The POWER switch controls power to the Power
Unit. The POWER INDICATOR light in the switch
will be lit when power is on.
3.2 TEMPERATURE SETTING
Push Buttons
When the Power Unit is turned on, the system
defaults to the LOW (90°F, 32°C) temperature
setting.
Temperature settings are selected by pressing the
appropriate push button-- MIN (fan only), LOW
(90°F, 32°C), MED (100°F, 38°C), HIGH (110°F,
43°C), or MAX (115°F, 46°C).
If electrical power is momentarily interrupted, the
existing temperature setting will remain. If
electrical power remains off, the TC3000 will
default to LOW (90°F, 32°C) when power is
restored.
Figure 4 - Operator Control Panel
Failure to follow the following
precautions could result in death or
serious injury:
• Do not use the MAX (115°F, 46°C)
setting in the operating room.
• Do not use the MAX (115°F, 46°C)
setting for patients with any of the
following conditions:
1. Low cardiac output
2. Poor peripheral perfusion or
peripheral vascular disease
3. Total immobilization
4. Non-sensate patients
3.3 REMOVE FROM USE Indicator
The REMOVE FROM USE indicator will light:
• if a patient limit thermostat or heater
overtemp thermostat has tripped, or
• if the HEPA filter has been installed
incorrectly or is missing.
The heating element and blower will automatically
turn off when either of these conditions occur.
If the REMOVE FROM USE indicator
lights, turn the Power Unit off
immediately. Discontinue use and refer
equipment for servicing.
®
WARNING
CAUTION
5