→ Keep the active electrode at least 15 cm (6 inches) away
→ DO NOT place the active cord or other equipment cables on
→ DO NOT allow the monopolar active cord or the cord to the
WARNING: DO NOT place dispersive electrode cables, active cords or
other electrosurgical accessory cables near cables leading to or from
cardiac or other patient monitoring equipment. DO NOT bundle cables or
cords with cables or cords from other electrical equipment as it may result
in interference.
WARNING: Capacitive Coupling. Wrapping electrosurgical instrument
active cords around metal objects can generate currents (capacitive
coupling) which may lead to electrical shock to the patient or user or fire.
WARNING: Electrical Shock Hazard. Follow the instructions for use
provided with the electrosurgical accessories for proper connection to the
electrosurgical generator. Prior to use, ensure all accessories are
properly connected to their receptacles and that no metal components of
the accessories are exposed as this may result in injury to the patient or
user.
WARNING: Electrical Shock Hazard. Wipe the active cord completely
dry before reuse.
Product Disposal:
When disposing of the product or its parts, ensure the proper disposal of hazardous and other
regulated waste in compliance with federal, state, and local regulations.
Issued Date: December 2019
Warning:
An issued or revision date for these instructions is included for the user's information. In the
event two years has elapsed between this date and product use, the user should contact
STERIS Endoscopy to determine if additional information is available.
Unless otherwise indicated, all marks denoted with ® or ™ are registered with the U.S. Patent and
Trademark Office, or trademarks owned by STERIS Corporation
Serious incidents that have occurred in relation to this medical device should be reported to the
manufacturer and competent authority in the country where the incident occurred.
US Endoscopy, a wholly owned subsidiary of STERIS Corporation.
Made in the U.S.A.
00732676B
from any monitoring electrode.
the patient.
patient return electrode to lie alongside, or be entwined with,
the cables to the monitoring electrodes.
Page 7 of 89