Guidance And Manufacturer's Declaration -Electromagnetic Emissions - CAIRE VisionAire Family Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 68
VisionAire Family
EMC Testing
Medical equipment needs special precautions regarding EMC and needs to be installed and put into ser-
vice according to the EMC information provided in this section.
Guidance and Manufacturer's Declaration —Electromagnetic Emissions
The VisionAire is intended for use in the electromagnetic environment specified below. The customer or the
user of the VisionAire should assure that it is used in such an environment.
Emissions Test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Recommended separation distances between portable and mobile RF
communications equipment and the VisionAire Units
The VisionAire is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of the VisionAire can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment (transmitters)
and the VisionAire as recommended below, according to the maximum output power of the communications
equipment.
Rated maximum
output power of
transmitter
from 150 kHz to 80 MHz
W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d
in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is
the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
20 - ENG
PN MN238-C4 C User Manual
Compliance
Electromagnetic Environment - Guidance
The VisionAire uses RF energy only for its internal function.
Group 1
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
Class B
The VisionAire is suitable for use in all establishments, including
domestic establishments and those directly connected to the
Complies
public low-voltage power supply network that supplies build-
ings used for domestic purposes.
Complies
Separation distance according to frequency of transmitter
d
1
2 .
P
0.12
0.38
1.2
3.8
12
m
from 80 MHz to 800 MHz
d
1
2 .
P
0.12
0.38
1.2
3.8
12
from 800 MHz to 2,5 GHz
d
1
2 .
P
0.23
0.73
2.3
7.3
23

Publicidad

Tabla de contenido
loading

Tabla de contenido