The Manufacturers or their agents overseas will provide a repair service for any unit developing a
fault.
Cleaning
If the instrument requires cleaning use a cloth that is only lightly dampened with water or a mild
detergent.
WARNING! TO AVOID ELECTRIC SHOCK, OR DAMAGE TO THE INSTRUMENT, NEVER ALLOW
WATER TO GET INSIDE THE CASE. TO AVOID DAMAGE TO THE CASE NEVER CLEAN WITH
SOLVENTS.
Tip Insulation
If the insulation of the probe tip becomes worn or damaged so that the inner layer of insulation shows
through, contact the Manufacturers or their agents overseas.
Calibration
The fundamental calibration parameter is magnetic field; calibration requires a calibrated field from a
precision Helmholtz coil. The manufacturer can provide a suitable re-calibration service in the event
that the local calibration service is unable to do so.
However, if only a current-in-wire re-calibration (with the toroid attachment) is required, this can be
achieved using a more widely available standard DMM calibrator. Calibration is performed using a 50
or 60Hz sine wave. Take care to establish a repeatable position of the cable within the opening in the
toroid; this is normally opposite the probe tip and as far back as possible.
Web link
For support please visit http://www.aimtti.com/support.