General specifications apply for the temperature range 5°C to 40°C with the probe connected to a
measuring instrument (oscilloscope or DMM) having an input impedance of 1MΩ || <30pF. Accuracy
specifications exclude errors of that measuring instrument and apply for the temperature range 18°C
to 28°C after 30 minutes warm-up and calibration at 23°C. Specifications quoted without limits are
typical characteristics determined by design and are not guaranteed.
Dynamic Characteristics
Bandwidth (small signal):
Pulse Rise-time (10% - 90%):
Propagation delay:
Pulse aberrations:
Slew Rate (equivalent):
Noise (equivalent in toroid):
Filter settings:
Magnetic Field Measurement
Scaling factor:
Accuracy and linearity:
Maximum field:
Current measurement in wire (with toroid attachment)
Current range:
Accuracy and linearity:
Scaling factor:
Maximum wire diameter:
Current measurement in PCB track
Scaling factor (with control
adjusted to suit track width):
General characteristics
Maximum output voltage:
Maximum bare-wire voltage:
Maximum track temperature:
Probe dimensions:
Cable length:
Safety & EMC:
DC to 5MHz.
<70ns.
60ns typical.
<±5% (<1% at lower bandwidth settings).
15A/µs
6mA
; 1.5mA
at minimum bandwidth setting.
rms
rms
Full bandwidth, 500kHz or 2Hz.
250µT (or 200A/m) per output Volt
±3%
±2·5mT (2000A/m)
±10mA to ±10A (DC + peak).
±5%
1 Amp per output Volt.
3.5mm (unbroken) or 6mm (end fed).
1 Amp per Volt for track widths 0·2mm to 3.5mm (0·007" to 0·14")
or 2 Amp per Volt for track widths 3mm to 6.5mm (0·125" to 0·25")
±10V, corresponding to ±2·5mT (field measurement) or ±10A (wire)
300V
CAT II
(circuits connected directly to the low voltage mains)
rms
600V
CAT I
(circuits not connected directly to the low voltage mains).
rms
Maximum Permitted Transient Overvoltage: 2500V.
The maximum temperature of a surface on which the probe
measurement tip can be placed for short periods (2 mins max) is
150ºC
155mm x 38mm x 28mm max; 2.8mm x 1.8mm at tip
2m from probe tip to output BNC
Complies with EN60950-1 & EN61326-1.
For details, request the EU Declaration of Conformity for this
instrument via http://www.aimtti.com/support (serial no. needed).