4. Operation
4.2. Setting-up
Connect the si629 P to mains. At power up the first time only the following screens will appear:
Instrument is already measuring pH, with the following factory settings::
• Control: ON/OFF
• Relay 2: cleaning during 3 sec. each hour, with a 5 sec. of post-cleaning time.
• Limit 1: input at pH 6.0, output at pH 6.5.
• Alarm: activation after 10 min of reading out of limits.
• Current output: 4 mA = pH 0, 20 mA = pH 14.
• Hold: fixed at 21 mA.
• Data logger: stores 1 datum each 30 min. (It is necessary to mark START, see page 27).
To modify this configuration see pages 17, 18 and 27.