Trouble Shooting - Xylem SI Analytics ViscoPump II Manual De Instrucciones

Unidad de medicion de la viscosidad
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Initial operation

2.5 Trouble shooting

Check whether the Basic Measuring Device is switched on:
Trouble
Air bubbles in
viscometer
Excessive pumping of
measuring medium
into the thermostat
bath or into the
safety bottle:
34
Trouble shooting
• Is the filling quantity sufficient? Check, fill viscometer if required.
• Is the viscometer of properly connected?
− in the case of pressing operation, please checked whether the filling
tube is connected; if necessary connect properly.
− for operation in suction mode, please check whether the capillary
tube is connected; if necessary, connect properly.
− please check whether the venting port is tightly connected; if
necessary re-tighten its screwed connection.
• Is the viscometer properly connected?
− for pressing operation?
− for operation in suction mode?
When using AVS measuring tripods:
• check position of the rack within the stand
• check the electrical connection from the viscometer to the ViscoPump
type II module
• Is the green LED on the measuring stand illuminated?
When using TC viskometers:
• Is the viscometer properly connected?
ViscoPump II

Publicidad

Tabla de contenido
loading

Tabla de contenido