For Use with CO
(Turn OFF the Air/Pump on the Light Source of the Processor)
2
1.
Prior to clinical use, inspect and familiarize yourself with the device and review the diagram below. If there is evidence of damage,
do not use this product and contact your local Product Specialist.
2.
Both Pinch Clamps should be in the open position when removed from the package.
3.
Open a bottle of sterile water and remove cap. (250cc, 500cc, or 1000cc).
4.
Remove 1/3 of the sterile water from the full sterile water bottle. This will allow for improved water pressure for better insufflation
and water flow.
5.
Open the sterile pouch and remove the AquaShield
contamination.
6.
Insert the Water Bottle Tubing into the sterile water bottle and firmly tighten the threaded Lid Cap to ensure a seal.
7.
Attach the Endoscope Connector to the endoscope.
8.
Leave the Blue Pinch Clamp on the Endoscope Tubing in the open position.
9.
Connect the CO
Connection Port/Luer Lock Connection to the luer lock connection on the CO
2
and the insufflator.
10. Leave the White Pinch Clamp on the CO
tubing is attached to a CO2 source/insufflator.
11. Follow the manufacturer's guidelines for the use of CO
12. Turn on the light source of the processor. Turn OFF the Air/Pump when using CO
13. Prime the air/water channel prior to insertion of the endoscope into a patient. If water pressure is low, check to see that the
Lid Cap is tightly closed. Ensure that all connections are secure.
14. Perform CO
insufflation and water flow for lens cleaning as described in the Olympus instruction manuals for air/water valve and
2
the endoscope being used together with a sterile water bottle.
15. If the water bottle needs to be replaced, use proper aseptic technique.
Note: If air is to be resumed, turn OFF CO
the White Pinch Clamp on the CO
source/insufflator.
After Procedure is completed
Follow the Olympus manufacturer's guidelines for purging the air/water channel system when the procedure is completed
Close the Blue Pinch Clamp on the Endoscope Tubing and disconnect from the endoscope. This will prevent water from
leaking from the Endoscope Connector.
Follow the manufacturer's guidelines for turning off the light source, the Air/Pump on the processor and CO
the procedure is completed.
When procedures are completed for the day, close the White Pinch Clamp and disconnect the CO
Connection from the insufflator.
Ensure that the device is discarded within 24 hours after the package has been opened. A label is provided with the device.
00731966 Rev. G
tubing in the open position when CO
2
source/insufflator and turn ON the Air/Pump on light source of the processor. Close
2
tubing if air is being used for insufflation and the system is not attached to a CO
2
Blue
Pinch Clamp
Endoscope connector
Lid Cap
Water Bottle
Tubing
®
system CO
. Avoid contact with the Water Bottle Tubing to reduce the risk of
2
source/insufflator.
2
White Pinch
Clamp
0.2 (max) micron
Hydrophobic Filter
CO
Connection Port/ Luer
2
Lock Connection
insufflator. Turn on the CO
2
is being used for insufflation and the
2
2.
source/insufflator when
2
Connection Port/Luer Lock
2
source
2
2
Page 3 of 34