The ME system is intended for use in the electromagnetic environment specified below.
The customer or the user of ME should assure that it is used in such an environment.
Emissions.test
Emissions
CISPR.14-1
Harmonic.emissions
IEC.61000-3-2
Voltage.fluctuations/Flicker.
emissions.IEC.61000-3-3
COMTENS III _3 languages.indd 21
Declaration – electromagnetic emissions
Compliance
Passed
Passed
Passed
21
Electromagnetic.environment.-.guidance
The.ME.system.must.emit.electromagnetic.energy.
in.order.to.perform.its.intended.function..Nearby.
electronic.equipment.may.be.affected.
The.ME.system.must.be.used.only.in.a.shielded.
location.with.a.minimum.RF.shielding.effectiveness.
and,.for.each.cable.that.exits.the.shielded.location,.
a.minimum.RF.filter.attenuation.of.80.dB.from.
10.MHz.to.20.MHz,.100.dB.from.20.MHz.to.80.
MHz.and.80.dB.from.80.MHz.to.100.MHz..(The.
minimum.at.20.MHz.is.100.dB.and.the.minimum.at.
80.MHz.is.80.dB.)
The.ME.system,.when.installed.in.such.a.shielded.
location,.is.suitable.for.use.in.all.establishments.
other.than.domestic.and.those.directly.connected.
to.the.public.low-voltage.power.supply.network.that.
supplies.buildings.used.for.domestic.purposes.
English