4.4.3 MAGNETIC CARD MODE Continued
Place the cuvette, containing a
stirring bar and the correct
volume of sample, into the
cuvette chamber.
See Appendix 1.
Supplementary
Air gap
Aspirate reagent
The addition of reagents
automatically starts the assay
cycle, which begins with a
blanking measurement.
Once the blank measurement
has been taken, the display
will show the time remaining
before the assay is complete.
At assay completion the result
will be displayed with the
appropriate units. The alarm
(beep) will sound, if switched
on.
Press ENTER to clear the
result from the display and
remove the used cuvette.
To perform a different assay,
press ESC and repeat the
process from the "Enter
Chemistry Number" stage.
Insert Code: EIN023
Place cuvette in
chamber
Wait 90 seconds for the
cuvette temperature to
stabilise
Press the blue button on
the pipette to carry out
the aspiration and
dispensing steps
Reagent sequence for
MININEPH assays
Reagent sequence for
MININEPH
assays
PLUS
Add reagent
Blanking
Assay time
XXs
Result =
XXXXXX g/L
CANCEL
Remove cuvette
Enter sample ID
ESC
Page 16 of 121
Temperature
stabilising
Aspirate reagent
Air gap
Supplementary
Upon dispensing the
reagents immediately
replace the pipette in the
pipette holder.
!
Samples suspected of being
in antigen excess will be
flagged „XS‟.
!
To perform the same assay
on further samples, repeat
the process from the "Enter
Sample ID" stage.
If no more assays are required,
switch off the MININEPH
Version: 11
.
PLUS
th
August 2011