EN-2
Replace Probe Cap Instructions
30253566
For optimum performance, use the following
operating procedures:
1. Take off the old probe cap: One hand
holds probe body, the other hand slightly
turns probe cap counterclockwise until it's
completely off from probe body.
2. Check whether O-ring is damaged:
Please remove the old O-ring if there is
any damage and replace a new O-ring.
3. Please make sure the new O-ring is
clean. Please use clean cloth to wipe out
any dusts or stains.
4. Put on new probe cap: one hand holds
the probe body, the other hand gently
turn the probe cap clockwise until the cap
is tightly secured.
5. Connect STDO21 to ST400D meter.
6. Input the probe cap parameter
To enter the parameter input setting, you
need to press READ + CAL + Setup button
together when "
selected.
STDO21 electrode
DO probe cap, STDO21, 1 unit
7. New Probe Cap Parameter
" is