12
Voltage detection, non-contact
(AC warning)
The non-contact voltage detector
integrated in the meter is able to
detect AC voltages from 100 V to
600 V. Switch the device on and run
the voltage sensor along the object
to be tested (5 - 10 mm). The display
(9) lights up if AC voltage is detected.
Non-contact voltage detection is no substitute for
!
conventional voltage testing. As the device detects an
electrical eld, it will react even to static charge.
Voltage detection, single-pole phase test
Remove the black test prod from the device. Turn the rotary
switch to position „V AC/DC" and connect the red test prod
to the phase or neutral conductor. The red display (9) only
lights up if the phase conductor is live. When the single-pole
phase test is carried out on the outer conductor, the indicator
function may be adversely affected under certain conditions
(e.g. when insulating personnel protective equipment is used
or at insulated locations).
The single-pole phase test is not suitable for checking
!
for zero voltage. To do this, you need to carry out a
two-pole phase test.
13
Flashlight function
To switch on the ashlight, press and hold down the
corresponding button. The light switches itself off
automatically as soon as the button is released.
14
Backlight
To switch the backlighting on, press the „Hold" button (11)
and hold it down for 2 seconds. Press the „Hold" button twice,
to switch the backlighting off.
15
Calibration
The meter needs to be calibrated and tested on a regular
basis to ensure it produces accurate measurement results. We
recommend carrying out calibration once a year.
MultiClamp-Meter Pro
19
GB