How To Use The Probe To Verify The Passivation Of Stainless Steel - Walter SURFOX Manual Del Usuario

Téster para pasivación del acero inoxidable
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

SURFOX Stainless steel passivation tester
giving you the possibility to enter another serial number. To change the value,
press 4 and enter the data using the keyboard.
Insert the SERIAL No and press 4, and position the probe on the test piece and
press 4 to begin testing.
After pressing 4 the test will start and a graphic bar will show the progression of
the test (Duration: 15 seconds).
After testing, the display shows the numeric value of the quality of the passive
layer of stainless steel between a maximum value (GOOD) and a "not passive"
(BAD).
Press 4 to return to SERIAL No to maintain the current session and the Project
No – or press the MENU button to end the session and return to the main menu.
NOTE: The tester has 50 memory blocks; once all the available cells are used,
to perform a new QC test, it will be necessary to clear the memory by following
the steps (see section 8.3.2 – Deleting Tests – User ID).
7. How to use the probe to verify the
passivation of stainless steel
1. Connect the probe to the external unit using the 4 pin cable supplied (Fig. 4).
2. Unscrew the tip protector and place it on the refilling cartridge compartment
cover (Fig. 5).
10
Fig. 4
Fig. 5

Publicidad

Tabla de contenido
loading

Tabla de contenido