Maintenance and troubleshooting
The accessories must not be repaired and handled with care.
Do not position the leg cuffs, arm cuff, abdominal belt
or other accessories near direct or indirect sources of
heat.
Do not place the device in damp or wet areas. Do not
soak the equipment in any way, do not pour any liquid
over the equipment.
Do not wash the equipment and accessories with water
or detergents. Simply wipe with a damp cloth.
Do not store the device with its accessories in a parti-
cularly cold area or under a temperature of 0° C. The
device could seriously be damaged.
If you don't use the device for a length of time, place
it back in its box.
Environment of use and safekeeping
• Use and maintain the device and its accessories in a clean environment and at a tempe-
rature between 10 – 40 °C, and humidity of 30 – 85%;
• Do not use in outside environments.
Problem
encountered
The unit does not turn
on
You hear strange hits
similar to that of esca-
ping air
The air does not flow to
the applicators, there is
insufficient pressure
The air continues to
flow into the chambers
and you hear an air
leak in the applicators
• Check that it is correctly connected to the mains;
• Check that the switch has been switch to ON (black ON/OFF
button) that is located on the side of the unit;
• Check the status of the power cable, it must be intact.
• Check to ensure that the hoses, the accessories and the con-
nector hoses are intact;
• Check to ensure that all the connections (do the unit and to the
applicators) have been inserted correctly;
• Check that the air connector hoses are not bent, squashed or
crushed.
• Check to ensure that the connector hoses to the unit and the
applicators have been correctly connected;
• Check that the air connector hoses are not bent, squashed or
crushed;
• Check that the chambers are inflated progressively and accor-
ding to the selected program;
• Make sure you set the pressure control correctly;
• Verify that the actual air outlet from the unit is present.
• Check for holes or damage in the applicators' air chambers. If
no abnormalities are present, switch off the device, turn it on
again to check again to ensure that is functioning properly.
52
User Manual: PressoSport
Possible solution
®