Introduction; Characteristics; Safety Notes - Festo DIDACTIC Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

GB

1. INTRODUCTION

These operating instructions contain user notes
for the Universal status display USD. This pre-
cision measuring device has a 4-digit LCD dis-
play and can be used universally.
D.C. voltage/current readings as well as basic
display of pressure, temperature and flow
values can be carried out using a wide range
of different sensors.
The user can adjust the device to suit different
kinds of sensors by means of simple pro-
gramming of additional sensor characteristic
curves.

2. CHARACTERISTICS

Compact device with 4-digit LCD display
Reading of pressure, temperature, flow rate,
voltage and current
Can be directly connected to almost all cur-
rent types of sensors with current or voltage
output, as well as Pt100 resistance ther-
mometers and NiCr-Ni thermocouples
MIN and MAX memory
Automatic value display
Multimeter function with autorange
Easy to handle
Can be adapted to a wide variety of sensors
through simple programming of additional
characteristic curves

3. SAFETY NOTES

These operating instructions contain informa-
tion and notes regarding the safe use and pro-
tection of the device, which must be strictly ob-
served.
Caution
To avoid electric shocks and damage to the in-
strument, do not connect voltages in excess of
30 V D.C. to the device.
Do not exceed the input values specified in the
table below, to avoid any damage to the de-
vice:
Input
V 0...30V, 0...10V
A 0...20mA
Pt100
NiCr-Ni
external voltage supply
Note
Do not connect voltage to input 'A 0...20mA
SP2' as this may cause damage to the device.
The input has been protected with a F0,1A in-
ternal fuse. If the current is too high, the fuse
will be blown and will require replacement. The
fuse must not exceed a nominal value of
100mA. Please observe the correct polarity
when using voltage input SP1 and current input
SP2.
- 12 -
maximum connected load
SP1
30VDC
100mA
SP2
(internally protected by fuse)
only resistance thermometers
SP3
Pt100 may be connected
only thermocouples of type
SP4
class K may be connected
24VDC +/-10%

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

183737

Tabla de contenido