MS 4
Put device switch (A) in the OFF position.
Plug in mains cable (G).
Commissioning
Once connected to the power supply the device is in "stand-by" mode.
Put device switch (A) in the ON position.
Any set values are retained when device is switched off and even after
Stirring
the device is disconnected from the power!
Set the engine speed with the operating button (E) on the right.
Heating
Contact thermome-
ter connection
Commissioning
MS 7
MS 10
HS 4
The right decimal point on the display (C) is lit.
Put device switch (A) in the ON position.
Any set values are retained when device is switched off and even after
the device is disconnected from the power!
Set the target temperature for the heating plate using operating button
(D) [for VHP (E)].
The set value is indicated on the display (C). If energy is being supplied
to the heating plate, the red LED (B) lights up.
In the stir and stand-by modes,
heat has been switched off for as long as the temperature of the set-up
surface exceeds 50°C.
Put device switch (A) in the OFF position, unplug contact
plug.
Safety contact thermometer acc. to DIN 12878 class 2 con-
nected with jack (H).
Put device switch (A) in the ON position.
Beware the instruction manual of the contact thermometer.
Notice: Display (C) also indicates the target temperature of
the heating plate when the contact thermometer is connec-
ted.
HS 7
HS 10
HP 4
flashes on the display after the
HP 7
HP 10
15