Use
To set the gas cylinder level indicator:
NOTICE
i
i
The gas cylinder level indicator is only available for
cylinders that are configured via a Dräger part no.
1. Select Gas level monit.
2. Activate or deactivate check box Gas level monit.
3. Confirm selection with OK.
To reset the gas cylinder level indicator for a new test gas
cylinder:
1. Connect a new test gas cylinder to a test gas connection.
2. Select
> Station gas configuration.
3. Select desired gas inlet.
4. Select Further options and select
cylinder level indicator.
26
5
!
i
i
to reset the gas
5.1
A visual inspection of the instruments must be conducted
every time before being inserted into the station.
1. Check that the housing, external filters and the nameplates
are intact.
2. Check and, if required, clean the housing, alarm elements
(LED window, horn opening), sensor ports and, if
applicable, battery contacts for contamination and damage
or have them checked by the service of Dräger.
i
i
Use
WARNING
A defective pressure reducer on the gas cylinder can
lead to increased pressure in the station. The gas
hoses may loosen as a result and gas may escape.
Health hazard! Test gas must not be inhaled. Observe
the hazard warnings in the relevant Safety Data
Sheets.
Provide venting into a exhaust or
outside the building.
NOTICE
To prevent loss of gas, Dräger recommends closing
the gas cylinders when the station is left unattended for
long periods.
Adjustment may not be possible due to instrument and
channel errors.
When using a compressed air cylinder at the
compressed air inlet, Dräger recommends to always
equip all modules with gas detectors prior to starting
the test. Otherwise, the compressed air cylinder will be
emptied unusually quickly.

Conducting a visual inspection

NOTICE
Instruments that did not pass visual inspection must
not be inserted into the station. Otherwise the test
cannot be correctly assessed in its entirety.
Dräger X-dock 5300/6300/6600