Use
If Test schedulermode is activated:
If necessary, log user out on the station (see chapter 4.5 on
page 23).
The preset test is performed according to the configured
time schedule.
Test passed:
Confirmation is shown on the display.
The Status LED flashes green.
If required, select the desired instrument field for additional
information (e.g. test date and performed tests).
Remove the instrument from the module.
WARNING
!
Prior to each use of the gas detector, check whether it
is switched on and in measuring mode after removing
it from the X-dock. Otherwise the user may carry a
device that is switched off!
Test passed, with reservations:
This state means that partial tests of the favourite could not be
conducted due to special settings.
Confirmation is shown on the display.
The status LED flashes yellow.
28
0 0
01033286.eps
0 0
01133286.eps
If required, select the desired instrument field for additional
information.
Remove the instrument from the module.
WARNING
!
Prior to each use of the gas detector, check whether it
is switched on and in measuring mode after removing
it from the X-dock. Otherwise the user may carry a
device that is switched off!
Test not passed:
An error message is shown on the display.
The Status LED flashes red.
If required, select the desired instrument field for additional
information.
Identify and rectify the error.
Repeat the test if necessary.
Overview of status LEDs
Colour
Status
blue
flashing
green
flashing
yellow
flashing
red
flashing
0 0
01133286.eps
Meaning
Process in progress
Test passed
Test passed,
with reservations:
Test failed/cancelled