Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

12. Specifications

12. Specifications
Specifications
Connections
EN
Inlet/return pumps
Outlet for gas cylinder
Outlet for gas recovery bag
Permissible pressure ranges
Inlet/return pumps
Outlet for gas cylinder
Outlet for gas recovery bag
TFT touchscreen
Display size
Resolution
Colours
Voltage supply
Battery power
Mains operated
Permissible temperature ranges
Operation
Storage
Flow measuring gas
Dimensions
Weight
SF₆ percentage sensor
Measuring principle
Measuring range
Accuracy
Resolution
36
SF₆-Q-Analyser
Quick coupling with self-sealing valve
Self-sealing valve DN8
Quick coupling, self-sealing valve
1.3 ... 35 bar abs. /1.3 ... 10 bar abs.
1.3 ... 10 bar abs.
< 1.015 bar abs.
7"
800 x 480
262,144
Lithium-ion battery, battery is charged during mains supply mode
AC 90 ... 264 V (50 ... 60 Hz)
0 ... 40 °C
-20 ... +60 °C
20 litres/hour
W x H x D: 538 x 406 x 297 mm
approx. 25 kg
Sound velocity
0 ... 100 %
±0.5 % based on SF₆/N₂ mixtures
(calibration for SF₆/CF₄ mixtures on
request)
0.1 %
WIKA operating instructions analytic instrument, model GA11
g³-Q-Analyser
1.3 ... 35 bar abs. /1.3 ... 12 bar abs.
1.3 ... 12 bar abs.
< 1.015 bar abs.
N/A

Publicidad

Tabla de contenido
loading