Before you start measuring, please calibrate the device. If you own multiple devices
make sure that the serial number on the back of the calibration plate matches the serial
number of your i1Pro device. Refer to the user guide of your measurement software for
additional information on the calibration process.
1. In the center of the calibration plate is a slider to protect the white calibration tile.
Open this protective slider until it clicks into place to access the white reference
ceramic tile on this calibration plate.
2. Place the i1Pro measurement device on the calibration plate. The device must seat
firmly on the calibration plate.
3. Set the software for calibration of the device and press the measurement button.
Status indicator lights switch off during calibration. Once the device is successfully
calibrated the
status indicator lights will pulsate red and then return to solid white.
4. When the calibration is finished you should close the protective slider for the white
reference ceramic tile.
A clean white reference ceramic tile is essential for providing accuracy of
your measurements. The protection cover should always be closed when the
white reference ceramic tile is not in use. If necessary you can clean the white
reference ceramic tile with isopropyl alcohol and a soft clean cloth.
status indicator lights pulsate white. If the calibration fails the