2x red flash
the row was not measured successfully because not all patches
could be recognized. Measure the row again, but reduce your
measurement speed and make sure that the device starts and
ends the measurement before and after the patches of the test
chart.
4x red flash
the row was not measured successfully because you started
reading the patches too early without giving the tungsten filament
lamp enough time to warm up. Measure the row again but allow
the lamp time to heat up before you start moving the i1Pro device.
1x green;
the row was measured successfully but the software
2x red flash
expected the measurement of a different row. Check if the row you
measured is the row the software is expecting.
5. Measure the same row again in the indicated direction. The second pass will
measure the row with the UV illumination only. If successful you can proceed to the
next row.
2x green flash the dual pass measurement of a row was successful.