Idiomas disponibles

5. Insert test strip into Test Port.
Meter turns on.
Note: If test strip has been out of the vial
too long before testing, an error
message appears upon insertion
of the test strip into the meter.
Release and discard old test strip.
Use new test strip for testing.
6. Wait until Drop Symbol appears in
Display. Keep test strip in meter until
testing is finished.
Note: If test strip is removed before testing is finished, an
error message appears. Release and discard old test
strip. Use new test strip for testing.
7. With cap removed, turn control
solution bottle upside down.
Squeeze one drop of control
solution onto a clean tissue. Wipe
off bottle tip and discard tissue.
8. Gently squeeze a drop of control
solution onto a small piece of
unused aluminum foil or clear
plastic wrap. Dispose after use.
18
Contact Blocks
Face Up
Drop Symbol
Discard Control
Solution Drop on
Tissue