Trouble Shooting Information - CareFusion MicroLoop Manual De Funcionamiento

Ocultar thumbs Ver también para MicroLoop:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 202

Trouble Shooting Information

Should you encounter problems operating your MicroLoop meter,
please consult the table below:
Problem
display 'freezes' and the
unit does not respond to
any key presses
No display present
Does not register a blow
Blows are inverted on the
display
Blows tracing ends
abruptly although patient
still exhaling
Battery does not hold
charge
Stylus does not register
icons on the display
Possible Cause
Multiple icons
have selected or
accidently
pressed
Charger not
connect or
battery is
exhausted
Head assembly or
cable broken
Head assembly or
cable broken
Turbine sticking
Exhausted
battery
Mains charger
faulty
Microloop not
seated in cradle
correctly
Touch screen
display requires
calibration
36
Solution
hold the on/off button
down for 10 seconds until
the unit switches off and
then turn on again
Connect charger to the
mains or return the unit
for servicing
Replacement of head
assembly or return the
unit for servicing
Replacement of head
assembly or return the
unit for servicing
Clean turbine in warm
soapy water or sterilising
solution, if problem
continues a replacement
turbine may be required
Replace the mains
charger
Re-position unit in the
cradle correctly, when
charging the orange light
will be illuminated.
Select the calibration
check icon and choose
touch screen, and follow
the instructions.

Publicidad

Tabla de contenido
loading

Tabla de contenido