Connecting Printers; Connecting The Line Voltage; Connecting A Network Peripheral - HEIDENHAIN QUADRA-CHEK 3000 Manual De Instrucciones

Ocultar thumbs Ver también para QUADRA-CHEK 3000:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 51
5.8

Connecting printers

Connecting a USB printer
Remove and save the dust protection caps
Route the cables depending on the mounting variant
Further information:
"Assembly of the device",
page 23
Connect the USB printer to one of the USB Type A
ports X31–X34. Make sure the USB cable connector is
fully inserted
Further information:
"Product overview", page 24
Pin layout of X31–X34
Further information:
"N", page 174
Connecting an Ethernet printer
Remove and save the dust protection caps
Route the cables depending on the mounting variant
Further information:
"Assembly of the device",
page 23
Connect the Ethernet printer to the Ethernet port X116
using a standard CAT.5 cable. The cable connector
must firmly engage in the port
Further information:
"Product overview", page 24
Pin layout of X116
Further information:
"O", page 174
5.9

Connecting a network peripheral

Connecting a network peripheral
Remove and save the dust protection caps
Route the cables depending on the mounting variant
Further information:
"Assembly of the device",
page 23
Connect the network peripheral to the Ethernet
port X116 using a standard CAT.5 cable. The cable
connector must firmly engage in the port
Further information:
"Product overview", page 24
Pin layout of X116
Further information:
"O", page 174
HEIDENHAIN | QUADRA-CHEK 3000 | Installation Instructions | 4/2016

5.10 Connecting the line voltage

WARNING
Risk of electric shock!
Improper grounding of electrical devices may result in
serious personal injury or death by electric shock.
Always use 3-wire power cables
Make sure the ground wire is correctly connected to
the ground of the building's electrical installations
WARNING
Fire hazard from the use of power cables that
do not meet the national requirements of the
respective country in which the product is
mounted.
Improper grounding of electrical devices may result in
serious personal injury or death by electric shock.
Use only a power cable that meets at least the
national requirements of the respective country in
which the product is mounted.
Use a power cable that meets the requirements to
connect the power connection to a 3-wire grounded
power outlet
Further information:
"Product overview", page 24
27

Publicidad

Tabla de contenido
loading

Tabla de contenido