About the Control Solution
• Only use the ACCU‑CHEK SmartView control solution.
• Close the control solution bottle tightly after use.
• Write the date you open the control solution bottle on the bottle label. The control solution must be
discarded 3 months from the date the bottle was opened (discard date) or on the Use By date on the
bottle label, whichever comes first.
• Do not use control solution that is past the Use By or discard date.
18
• Refer to the control solution package insert for control solution storage conditions.
• The meter automatically recognizes the difference between the control solution and blood.
• The control results are not displayed in memory.
• The control solution can stain fabric. Remove stains by washing with soap and water.