Table 1 List of part and spare parts
Pos.
20
30
40
50
52
60
70
80
90
n.c.
n.c.

4.1.3 Line Filter

Danger of shock hazard
For measurements in circuits with voltages V (RMS) > 30 V, suitable measures (e.g.
appropriate measuring equipment, fusing, current limiting, electrical separation,
insulation) should be taken to avoid any hazards.
As described in Fig. 3, the line filter must be connected to the transformer (see also Fig. 6).
When the power switch is activated and the 230 V supply voltage is present, 230 V AC
should be present between outputs A and D of the line filter. For 115 V supply voltage, 115
V should be present between outputs A and D and between outputs B and C. If this is not
the case, check the line filter and exchange it as required.
Fig. 6 Line filter
Make sure that the connections between the line filter and transformer are bound
together with a cable tie to eliminate the possibility of an electrical connection between
the primary and secondary sides of the transformer (see Fig. 6).
Manual 1316.4969.02 - 01
Designation
PRINTED BASE PLATE
REGULATOR BOARD
READY TO USE TRANSFORMER
LINE FILTER WITH VOLTAGE SELECTOR
CABLE TIE
PRINTED FRONT PANEL
INPUT CABLE W1
with 12-contact socket (see Fig. 4)
GROUNDING CABLE
COVER
DEVICE PEDESTAL TRANSPARENT
FUSE IEC 60127-2/3 T125mA H 250V
(mounted in line filter)
Stock no.
1316.4968.00
1303.0317.02
1303.0369.00
1120.8804.00
0015.9038.00
1316.4951.00
1303.0330.00
1303.0375.00
1316.4980.00
3552.3610.00
0009.5528.00