Troubleshooting
If there is a problem, try the following suggestions. If the problem cannot be solved, contact your
equipment supplier or ResMed. Do not attempt to open these devices.
Problem/Possible cause
No display
Power is not connected.
The DC plug is partially inserted into the
back of the device.
The S9 and H5i are not connected
correctly.
Insufficient air delivered from the device
The S9 is not working correctly.
Air tubing is not connected properly.
Air tubing is blocked, pinched or
punctured.
The H5i flip lid is not latched correctly.
The H5i flip lid seal is not fitted
correctly.
Incorrect air tubing selected.
Displays error message: Heated tube fault, replace tube
Device has been left in a hot
environment.
There is a fault in your ClimateLine or
MAX
ClimateLine
air tubing.
Displays error message: Humidifier fault, replace humidifier
There is a fault in your H5i heated
humidifier.
Solution
Ensure the power cable is connected and the power outlet
(if available) is on.
Fully insert the DC plug.
Ensure that the S9 and H5i are securely attached.
See the Troubleshooting section of the S9 VPAP Tx Clinical
Guide.
Check air tubing.
Unblock or free the air tubing. Check the air tubing for
punctures.
Close the flip lid ensuring that it clicks into place.
Make sure the flip lid seal is facing the right way up and
fitted securely.
If you are using the SlimLine or Standard air tubing ensure
that you have the correct air tubing selected via the menu.
Allow to cool before re-use. Disconnect the power cord
and then reconnect it to restart the device.
Discontinue using your ClimateLine or ClimateLine
tubing and contact your service provider. Use Standard or
SlimLine tubing in the interim.
Discontinue using your H5i and contact your service
provider.
MAX
air
5
English