Constant Conductance Mode (CG)
Conductance Ranges:
Setting Accuracy:
Regulation:
Temperature Coefficient:
Slew Rate Ranges:
Minimum transition time:
Constant Voltage Mode (CV)
Voltage Ranges:
Setting Accuracy:
Regulation:
Temperature Coefficient:
Slew Rate Ranges:
Minimum transition time:
Constant Power Mode (CP)
Power Range:
Setting Accuracy:
Regulation:
Temperature Coefficient:
Slew Rate Ranges:
Minimum transition time:
TRANSIENT CONTROL
Transient Generator
Pulse Repetition Rate:
Pulse Duty Cycle:
Setting Accuracy:
Slew Rate Control
The slew rate control applies to all changes of level whether caused by manual selection, remote
control or the transient generator.
The level change is a linear slew between the two level settings with a slope determined by a
three-position range switch and a single turn control. Read-out is in the units of the mode
selected; the range available in each mode is shown above.
Setting Accuracy:
Level Variation:
4
<0·01 to 1 A/V (1 mA/V resolution); <0·2 to 40 A/V (0·01 A/V
resolution).
± (0·5% + 2 digits + 30 mA).
< 2% for 90% load power change (V > 2 Volts, using remote sense).
< (0·04% + 5 mA) per ºC.
1 A/V range: <0·1 A/V per s to >10 A/V per ms.
40 A/V range: <4 A/V per s to >400 A/V per ms.
100 µs.
V
to 8 V (1 mV resolution) and V
min
V
depends on current: typically <2V at 80A
min
± (0·2% + 2 digits).
< 30 mV for 90% load power change (using remote sense).
< (0·02% + 1 mV) per ºC.
8 V range:
<0·8 V per s to >80 V per ms.
80 V range:
<8 V per s to >800 V per ms.
100 µs.
0 to 320 Watts.
± (0·5% + 2 W).
< 2% over 5 V to 75 V source voltage change (using remote sense).
< 0·1% per ºC.
<32 W per s to >3200 W ms.
100 µs.
Adjustable from 0·1Hz to 10kHz in 3 overlapping ranges.
1% to 99% (percentage of period at Level B).
Frequency: ± 2% of range; Duty Cycle: +2%
± 10% ± minimum transition time (levels within 10% to 90% or range).
± 5 digits
to 80 V (10 mV resolution).
min