Methods; Chlorine, Low Range (Cl 2) - Orbeco SC450 Manual De Instrucciones

Ocultar thumbs Ver también para SC450:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Methods

CL 2
Tablet
Powder
0.0.0
Tablet
Powder
Zero
Test
CL 2
Tablet
Powder
RESULT
Tablet
Powder
Zero
!
Test
CL 2
Tablet
Powder
RESULT
Tablet
Powder
8
Chlorine with Powder Pack
0.02 – 2.0 mg/l
a) free Chlorine
Fill a clean vial (24 mm Ø) with 10 ml of the water sample and
perform zero calibration (see "Operation").
Remove the vial from the sample chamber.
Add the contents of one Chlorine Free-DPD/ F10 Powder Pack
straight from the foil into the water sample.
Close the vial tightly with the cap and invert several times to mix the
contents (20 sec.).
Place the vial in the sample chamber making sure that the
are aligned.
Press the [ZERO/TEST] key.
The method symbol flashes for approx. 3 seconds.
The result is shown in the display in mg/l free Chlorine.
b) total Chlorine
Remove the vial from the sample chamber.
Rinse the vial and the cap several times and then fill the vial with
10 ml of water sample (Note 2 and 3).
Add the contents of one Chlorine Total-DPD/ F10 Powder Pack
straight from the foil into the water sample.
Close the vial tightly with the cap and invert several times to mix the
contents (20 sec.).
Place the vial in the sample chamber making sure that the
are aligned.
Wait for a reaction period of 3 minutes.
(Countdown can be activated, see page 23)
The method symbol flashes for approx. 3 seconds.
The result is shown in the display in mg/l total Chlorine.
c) combined Chlorine
combined Chlorine = total Chlorine – free Chlorine
Tolerances:
0 – 1 mg/l: ± 0.05 mg/l
> 1 – 2 mg/l: ± 0.10 mg/l
marks
marks
SC450_V1_11.2010

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Hellige sc450

Tabla de contenido