Lovibond Tintometer MD200 Manual De Instrucciones página 43

Ocultar thumbs Ver también para Tintometer MD200:
Tabla de contenido

Publicidad

Methods
GB
Notes:
1.
Vial cleaning:
As many household cleaners (e.g. dishwasher detergent) contain reducing substances,
the subsequent determination of Bromine may show lower results. To avoid any meas-
urement errors, only use glassware free of Chlorine demand.
Preparation: Put all applicable glassware into Sodium hypochlorite solution (0.1 g/l) for
one hour, then rinse all glassware thoroughly with deionised water.
2.
Preparing the sample:
When preparing the sample, the lost of Bromine, e.g. by pipetting or shaking, must be
avoided. The analysis must take place immediately after taking the sample.
3.
The DPD colour development is carried out at a pH value of 6.2 to 6.5. The reagents
therefore contain a buffer for the pH adjustment.
Strong alkaline or acidic water samples must be adjusted between pH 6 and pH 7 before
the reagent is added (use 0.5 mol/l Sulfuric acid resp. 1 mol/l Sodium hydroxide).
4.
Exceeding the measuring range:
Concentrations above 22 mg/l Bromine can lead to results showing 0 mg/l. In this case,
the water sample must be diluted with water free of Chlorine and the measurement
repeated.
5.
Oxidising agents such as Bromine, Ozone etc. interfere as they react in the same way as
Bromine.
Reagent
DPD No. 1
MD200_3 02/2014
Form of reagent/Quantity
Tablet / 100
Order-No.
511050BT
43

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Md200

Tabla de contenido