Separation Distances

Recommended Separation Distances Between Portable And Mobile RF Communications Equipment
The EntraFlo™ Feeding Pump is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer
or the user of the EntraFlo™ Feeding Pump can help prevent electromagnetic interference by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitters) and the EntraFlo™ Feeding Pump as recommended below, according to the maximum output
power of the communications equipment.
Rated Maximum
Output Of Transmitter W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above the recommended separation distance d in metres (m) can be estimated using
the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according
to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.

Specifications

TYPE OF DEVICE:
Volumetric enteral feeding pump
And The EntraFlo™ Feeding Pump
Separation Distance According To Frequency Of Transmitter M
150 kHz to 80 MHz
d = 1.17√P
0.12
0.37
1.17
3.69
11.67
80 MHz to 800 MHz
d = 0.35√P
0.04
0.11
0.35
1.11
3.50
MECHANISM:
Rotary Peristaltic
800 MHz to 2.5 GHz
d = 0.7√P
0.07
0.22
0.70
2.21
7.00