Recommended Separation Distances Between Portable And Mobile RF Communications Equipment
The COMPAT
Standard Enteral Feeding Pump is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The
®
customer or the user of the COMPAT
between portable and mobile RF communications equipment (transmitters) and the COMPAT
according to the maximum output power of the communications equipment.
Rated Maximum
Output Of Transmitter W
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above the recommended separation distance d in metres (m) can be estimated using the equation
applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures,
objects and people.
Specifications
TYPE OF DEVICE:
Volumetric enteral feeding pump
And The COMPAT
Standard Enteral Feeding Pump
®
Standard Enteral Feeding Pump can help prevent electromagnetic interference by maintaining a minimum distance
®
Separation Distance According To Frequency Of Transmitter M
150 kHz to 80 MHz
d = 1,17√P
0,12
0,37
1,17
3,69
11,67
Standard Enteral Feeding Pump as recommended below,