Co Montior Troubleshooting guide
Monitor will not calibrate at 10 ppm.
Monitor will not successfully complete a zero point
adjustment.
Monitor displays a (-0) reading or rC.
Monitor zeroes during the zero point adjustment and the
reading elevates after returning to the rUn mode.
Monitor will not turn On.
Monitor does not display flashing green light after calibration
or a partial display appears on the monitor screen.
Monitor does not turn OFF.
Monitor displays an (F1) reading.
Monitor displays (8.88, 2F or 9E) with flashing lights.
Monitor shuts down after 60 second countdown.
Monitor displays "lb".
Yellow caution LED flashes.
Americas:
E.D. Bullard Company
1898 Safety Way
Cynthiana, KY 41031-9303
Toll free: 877-BULLARD (285-5273)
Tel: 859-234-6616
Fax: 859-234-8987
www.bullard.com
Europe:
Bullard gmbH
Lilienthalstrasse 12
53424 Remagen • Germany
Tel: +49-2642 999980
Fax : +49-2642 9999829
www.bullardextrem.com
1. Prior to performing a calibration, allow the temperature of the monitor to
stabilize to the environment in which it will be used for several minutes.
2. The Clean Air Box should be under pressure with air flowing to the box with
no air going through the outlet couplers (i.e. under static pressure).
3. Perform a zero point adjustment using impurity-free air.
4. If the instrument zeroes, perform a calibration using 10 ppm CO.
5. Verify that the calibration gas is 10 ppm and the gas cylinder is not empty. (Gauges are available.)
6. If the instrument will not calibrate, remove the air supply and repeat both
the zero point adjustment and the 10 ppm CO calibration.
7. If the instrument will not calibrate, replace the sensor and repeat both the zero point adjustment and the 10
ppm CO calibration.
• Wait 5 minutes and repeat steps 1-4 above.
• Wait 5 minutes and repeat steps 1-6 above.
A. Allow unit to stabilize in the run mode for 5 minutes.
B. Repeat steps 1-6 above.
C. If the unit will not calibrate, the supply air may be contaminated.
D. Check the supply air for CO level.
• Verify that the ON/OFF button under the front cover of the monitor has
been pressed to activate the monitor
• Check AC power supply
1. Confirm the ON/OFF switch on the control box is in the ON position
2. Confirm AC power cord is connected to working outlet (Portable)
3. Confirm unit is wired correctly (Panel mount)
• Using Battery back up
1. Confirm the AA batteries are good – replace if needed
• Using 9V battery installed in monitor
1. Verify that the battery is good – replace if needed
• If the power supply is good and the monitor will not power up, return it to the factory for repair or replacement.
• With the monitor ON and the unit on battery power only, remove and reinstall the 9 volt battery.
• Repeat steps 1-6 above.
• Monitor must be in the RUN orTEST mode.
•The rotary cam switch is not locked in the CAL,TEST or RUN position.
•Turn the switch knob slightly until it is in the correct position.
nOTE
1. AC appears on the display if attempting to go to CAL mode.
2. CO and an alternating number appears on the display if the unit
is in the TEST mode.
3. A number (PPM content of CO in the compressed air) appears
on the display if attempting to go to the RUN mode.
•This can occur when the ON / OFF button is pushed while the monitor is in
the fault mode with F1 showing on the display.
• Continue to push the ON / OFF button until F1 is shown on the display.
•Turn the switch knob slightly to lock the monitor into the CAL,TEST or RUN mode.
• Check power supply.
• Verify that the battery is good.
• If the power supply is good and the monitor will not power up, return it to the factory for repair or replacement.
This indicates a low battery condition and the 9V battery in the CO monitor
should be replaced.
•This is normal forTEST mode and no action is necessary.
• If the unit is not inTEST mode, then it may indicate a low battery condition.
Replace the 9V battery in the CO monitor.
Asia-Pacific:
Bullard Asia Pacific Pte. Ltd.
LHK Building
701, Sims Drive, #04-03
Singapore 387383
Tel: +65-6745-0556
Fax: +65-6745-5176
www.bullard.com
©2011 Bullard. All rights reserved.
Bullard is a registered trademark of Bullard.
It's your life and you're worth it is a trademark of Bullard.
6081257771D (0411)