Metcal PCT-1000 Guia Del Usuario página 5

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

prevent operators from editing profiles etc. To activate this function, a 4
digit number must be entered within a program setup as explained
below.If the pass code is set at 0000 the programs are not protected.
When C: 0000 appears on the screen, you can select a 4 digit pass
code.
7.
When the first digit flashes, use the  or ⊕ buttons to select the
preferred number.
8.
Press (↵) and repeat step 7 for all digits.
9.
Press-and-hold (↵) to quit code entry.
10.
From this point on you can only edit profiles after entering the
correct code
Security Lockout Notes:
You can 'turn off' the pass-code protection by first going into
Program mode, entering the current code then setting the new code
to '0000' – All users now have full access to all program functions.
One code protects all profiles, not individual profiles.
If the Pass-code is lost or forgotten then the Master Reset function
will restore the factory defaults including erasing the pass-code
10.
Press (↵) to move to the next parameter. Turbo mode (TM) is for
regions with <120VAC. Use ⊕ or  to switch between YES or NO or
press (↵) to skip
11.
Once required changes have been made, press and hold (↵) for 3
seconds until the LCD tops flashing, to set the profile to memory and
exit setup mode.
Master Reset.
To reset the unit it to factory settings, the following procedure can be used.
Press and hold ⊕ and (↵)
Whilst keeping ⊕ and (↵) depressed, switch the unit on
using 
PLEASE NOTE: Use of the master reset will erase all stored programs.
2.
Active Setup Mode
1.
Select Profile to be modified as per steps 4.1.1 to 4.1.4
2.
Once the temperature in Z1 has started to flash, Activate Setup Mode
by depressing the footswitch
3.
Profile will start to run and 'SETUP' will be displayed on LCD.
4.
To modify the current zone set-temperature use the ⊕ /  buttons.
5.
To increase the time of the current zone press and hold the (↵) button
as the zone ends. Instead of stepping onto the next zone the current
zone time-setting will increment the original setting for as long as you
keep the (↵) button depressed. Releasing the button will allow the
profile to progress onto the next Zone.*
6.
Note: Steps 4 & 5 can be carried out simultaneously.
7.
To decrease the zone time, press the footswitch at the desired time.
This will set the time to memory and advance the profile into the next
zone.*
8.
Once required changes have been made to zone 1-4, allow the profile
to finish the end of the cool down zone. Press and hold (↵), until the
LCD tops flashing, to set the profile to memory and exit setup mode.
*Note: feature not available in Cool-Down Zone
4

Publicidad

Tabla de contenido
loading

Productos relacionados para Metcal PCT-1000

Tabla de contenido