Recording the test data
• After the presets have been entered correctly, the device is ready to record.
• Press the
› key to start a test. The message «PLEASE WAIT USB»
and the file number appear flashing on the display (4).
Now the file is opened and the header created
• As soon as the file has been created, the symbol on the display
(4) changes from | to a flashing " and the test is started.
• The force, speed and expansion values are recorded every millimetre.
• A header with client text, file number, date, time and settings is saved for every test (see Evaluation).
Stopping the recording
• The test is stopped automatically if the sample tears (F
if it is cancelled with keys
the track is reached.
• The message «PLEASE WAIT USB» appears on the display again.
Never remove the stick while this message is still displayed!
Further recordings
• Take off the probe
• If the values are not recorded any more, the USB stick must be removed.
• From time to time, check the remaining memory of the USB stick on a PC.
› ,
,
or if the end of
+
‹‹
››
Return to Initial Length