7.2
Tescom Valve Maintenance
7.2.1
The following items are needed to complete valve repair:
2 ½ inch open end wrench (101582697)
2 inch open end wrench (101582698)
Repair Kit (101629337)
Torque Sensor Tool (101629332)
Loctite 271
A light lithium or silicone based grease
7.2.2
Front Panel Removal
7.2.2.1 Remove the cell assembly from the tower. The weight of the cell could
cause the instrument to tip when the drawer is opened.
7.2.2.2 Remove power and air from the iX77.
7.2.2.3 Ensure the water and drain lines have enough slack to move forward
without binding.
7.2.2.4 Remove the six screws from the sides of the front panel.
7.2.2.5 Pull the oil bottle drawer out.
7.2.2.6 Rotate the front panel gain access to the Tescom valve.
7.2.2.7 Locate the latch behind the Tescom valve.
101625591

Tools Needed

Figure 7-1
Rear Fluid Connections
iX77™ Rheometer User Manual
Revision F
104