EqUIPMENT'S OPERATION
Operating guidelines
The unit is designed with overheating protection, to avoid extended and continuous
operation for long time. When the red light is on, the device cannot be activated, even
if it is forced to do so. The unit remain idle for about 20 minutes and can be restarted
after the green light illuminates.
b. The degassing function can be used. Press the SET button to set the clock. Press the
functions button, it lights up
device switches to the normal cleaning mode after 90 seconds.
c. When the "smoke" of the waste is no longer visible, the cleanup is complete. If an
additional cleaning is required, press the SET button to reset the clock and repeat the
above steps.
18
Press the On/Off button once to start the
cleaning.
During cleaning, a humming can be heard coming
from the cleaning tank, indicating that cleaning is
in progress.
Closing the lid will reduce the noise level.
The LED display will count down, showing the
remaining cleaning time. When the display reads
0:00, the cleanup is complete. To interrupt the
cleaning at any time, press the On/Off button.
Select one of four cleaning methods previously
recommended.
a. When the heater is needed, press the TC button, it
flashes
, indicating that the heater is activated.
When the water temperature reaches 60ºC, the
heater will automatically shut down.
To ensure safety, the unit is designed with double
protection.
The heater will automatically shut down if running
for 45 minutes.
. Press the On/Off button to start the degassing. The
When cleaning is complete, disconnect the power
cord, remove the lid and the objects. Pour water
through the retreat. Clean and dry the tank.