IDENTIFIcATION OF THE EqUIPMENT

Principles and bases applied to the operation of the product

Cleaning occurs by cavitation, reaching uniformly throughout the material to be clean,
even in hard to reach places.
This occurs through an ultrasonic vibration generator, where a field test is subjected to
a vibration in high frequency when immersed in liquid solution.
The vibration generates a negative pressure causing the emergence of micro bubbles in
the liquid and a positive pressure providing the collapse of these bubbles on the surface of
the proof body forming micro jets that are responsible for erosion and cleaning effect.

Description of the Equipment

Ultrasound equipment for cleaning medical and dental instruments and accessories
composed of:
Injected body in ABS;
Control panel with digital display;
Transparent and removable viewer cover,
Tank made of stainless steel bowl and plastic basket;
Ultrasonic vibration generators;
Power cord.

Indication of Equipment

Medical, dental and laboratory instruments and supplies cleaning "removal of blood
and small particles on the instruments prior to disinfection". It should be used and handled
by qualified person (professional properly regulated, according to local legislation of the
country) following the instructions contained herein.
It is the user's obligation to use only the equipment in perfect conditions and protect
himself/herself, patients and others from possible dangers.
This equipment performs the cleaning of instruments and does not perform disinfection
and/or sterilization processes.