stimulator . These electrodes are commonly available in the hospital, and
must comply with the guide ISO 10993-1 on the compatibility, and are not
supplied with the stimulator .
4. TECHNICAL DATA
4.1
General information
Power supply:
Battery duration:
Impulse frequency:
Impulse duration:
Current impulse:
Quantity of charge:
Measure unit selection:
Display:
Operating temperature: Room temperature between 10°C and 40°C and
4.2
Output waveform
The output waveform has the typical course shown in the following figure,
where two cases with resistance of 1000 e 500 Ω are reported, with an impulse
duration of 300 µS and a current of 1,0 mA:
0.5
internal alkaline battery 9 V, 500 mAh (6LR61);
more than 25 hours of continuous functioning;
1 Hz, 2Hz, 4 Hz, selectable by the "µS/MODE"
button on the front panel .
50 µs, 100 µs, 300 µs, selectable by the "MS/MODE
" button on the front panel .
adjustable by the knob on the front panel or the
side control: from 0 . 0 mA to 6 . 0 mA for duration of
50 µS; from 0.0 mA to 5.0 mA for duration of 100
µS; from 0.0 mA to 4.0 mA for duration of 300 µS.
adjustable by the knob on the front panel or the
side control: from 0 to 300 nC for duration of 50 µS,
from 0 to 500 nC for duration of 100 µS, from 0 to
1200 nC for duration of 300 µS.
mA or nC, selectable by the menu .
LCD custom 4 digits and icons for the visualization
of the current, or the quantity of charge, and for the
set up of the stimulator .
relative humidity includes between 30% and 75% .
V
1
R=1000 Ohm
300
s
µ
R=500 Ohm
12
s
µ