Attention!
After adhesive bonding, the wedge cannot be removed any more
from the heel part.
Attention!
Do not force the wedge further into the toe springs.
Do not substitute other objects in place of recommended wedges.
7 Maintenance and service
The prosthetic foot is maintenance-free.
Generally, all moveable parts must be inspected visually and for functionality
every 6 months; all non-moveable structural parts every 12 months. If the
product is used for moderate sportive activities, the inspection intervals must
be reduced to a maximum of 3 months.
Note:
If necessary, replace foot shell and / or spectra-sock.
Permanent bond (final mounting)
Apply a thin line of Cyanoacrylate
adhesive around the bottom perimeter
of the wedge. When installing the
wedge, align the posterior edges of
wedge and pad. Press on the wedge
and hold it for 15 seconds.
21/40