CAUTION:
U.S. DEPARTMENT OF TRANSPORTATION (DOT) AND UNITED NATIONS (UN) REGULATIONS REQUIRE
THE REMOVAL OF BATTERY FROM THE DEVICE FOR ALL INTERNATIONAL AIRLINE TRAVEL WHEN
THE ECLIPSE IS CHECkED AS LUGGAGE.WHEN SHIPPING THE ECLIPSE, THE BATTERY MUST ALSO BE
REMOVED FROM THE DEVICE AND PACkAGED PROPERLY.
DO NOT ATTEMPT TO OPEN THE BATTERY, THERE ARE NO SERVICEABLE PARTS INSIDE THE BATTERY.
ONLY USE SEQUAL-PROVIDED BATTERIES.
RETURN THE BATTERY TO YOUR HOME CARE PROVIDER FOR PROPER DISPOSAL.
kEEP THE BATTERY AWAY FROM CHILDREN.
TypICAl bATTERy REChARGE TImE
the typical time to recharge your battery, in order to achieve 80% capacity, from a fully
discharged battery is up to 5.0 hours, depending upon the device flow setting. When the battery
is fully charged, the battery gauge will cease cascading.
if the battery becomes too warm during discharging, recharging will not begin until the battery
sufficiently cools. removing the battery and allowing it to cool may expedite this cooling process.
the battery may not recharge when operating from the DC Power Supply, depending on flow
setting and temperature.
the battery charges when in the device or with the optional Desktop Charger.
OpERATING yOUR EClIpSE FOR ThE FIRST TImE
Step 1: positioning your Eclipse for Use
Place the eclipse in a well-ventilated, well lit area. be sure the air inlet
and exhaust vent are not obstructed.
Position the eclipse so that all audible and visual indicators or alarms
can be easily seen and heard.
be sure the air inlet filter is in place before operating your eclipse. if
the air inlet filter is missing, contact your home care provider. if the
air inlet filter is dirty, wash with warm soapy water, rinse with clear
water and allow it to dry before placing the air inlet filter back in the
device. if necessary, replace it with a new, clean air inlet filter.
Plug the device into a grounded AC Power outlet, a DC Power
source, or be sure there is a fully charged battery installed.
Step 2: power ON the Device and Allow it to Warm-Up
Press and hold the "ON/OFF" button for two (2) seconds to power ON your eclipse. A brief
audible and visual power on self-test will occur. Verify that all the indicators are illuminated and
18
that a buzzer sounds indicating the device is operating properly.
Personal Ambulatory Oxygen System Users Manual
Air Inlet Filter
Exhaust vent