the air inlet filter should be replaced as needed. if the eclipse is used in a dusty environment,
the filter may need to be replaced more frequently. You should not operate the eclipse without
an air inlet filter installed.
AC power Supply, DC power Supply, power Cord and Outer Case: turn OFF the eclipse and
disconnect from AC or DC power before any cleaning or disinfection activity. DO NOT spray
the outer case. use a damp (not wet) cloth or sponge. Spray the cloth or sponge with a mild
detergent solution to clean the cabinet and power supplies. to disinfect the eclipse, use lysol
brand ii disinfectant (or equivalent). Proceed as directed by the manufacturer.
WARNING:
UNPLUG POWER CORDS, AC OR DC POWER SUPPLIES BEFORE CLEANING THE EXTERIOR CABINET.
DO NOT USE DENATURED ALCOHOL OR APPLY LIQUID SPRAY OR AEROSOL CLEANERS TO THE CORDS
OR POWER SUPPLIES.
Optional humidifier: if used, clean your humidifier daily to reduce the possibility of
contamination. Follow the cleaning recommendations provided by the home care provider, or
the manufacturer of the humidifier. replace the humidifier monthly, or as recommended by the
manufacturer, or your home care provider.
Universal Cart: before cleaning the universal Cart, remove it from the eclipse. use a damp
(not wet) cloth or sponge and mild detergent solution to clean the universal Cart.
power Cartridge (battery) Cleaning, Care and maintenance
the battery in the eclipse requires special care to assure a longer life and the highest level of
performance. the SeQual battery is the only approved battery recommended for use with the
eclipse.
use a damp (not wet) cloth or sponge to clean the battery. First spray the cloth or sponge with
a mild detergent then clean the battery case and the latch.
battery Disposal: Your battery is rechargeable and can be recycled. Always return it to your
home care provider for proper disposal. You can also contact your local city or town offices for
instructions on proper disposal of the battery.
WARNING:
USE OF A DAMAGED BATTERY MAY CAUSE PERSONAL INjURY.
PROTECT THE ECLIPSE AND ALL POWER SUPPLIES FROM FLUID SPILLS OR FLUID DRIPS TO AVOID
POSSIBLE SHOCk HAZARDS.