WARNING:
A BACk-UP SOURCE OF SUPPLEMENTAL OXYGEN IS RECOMMENDED IN CASE OF A POWER OUTAGE
OR A MECHANICAL FAILURE OF THE DEVICE. CONSULT YOUR PROVIDER FOR A BACk-UP SOURCE
OF OXYGEN.

TROUblEShOOTING TAblE

SympTOm
Eclipse does not
power on when
ON/OFF button
is pressed
No Oxygen
Low Oxygen
Concentration
Low Oxygen Flow
No Oxygen
Delivered in
Pulse Flow Mode
Power Cartridge Status
Gauge never indicates
22
fully charged
Personal Ambulatory Oxygen System Users Manual
DO NOT IGNORE AlARmS.
pRObAblE CAUSE
No Battery installed
Battery is discharged
or no external power is present
AC or DC Power Supply
is not plugged in correctly
Insufficient power from the outlet
Other
Filter Blocked
Humidifier
Eclipse not ON
Tubing or cannula is not properly
connected or kinked
Other
Restriction in tubing or humidifier
Filter restricted
Exceeding maximum breath rate in
Pulse Mode. *See Pulse Dose Mode
Breath Rate Chart (p. 18)
Other
Restriction in humidifier or tubing
Filter Blocked
Other
Tubing/cannula longer
than 7 feet (2.1m)
Humidifier attached
No inspiration detected
Unit in Continuous Mode
Nasal cannula not in the
nose or mouth breathing
Power Cartridge (Battery) is aging
REmEDy
Install
or plug into external power
Battery
Plug into external power
Check plug at outlet, Power Supply box and
at the Eclipse to ensure properly plugged in.
Ensure there is adequate power from the outlet.
(Outlet may be controlled by a wall switch
or the circuit breaker may have engaged.)
Contact your dealer/provider
Clean Air Inlet Filter
Check humidifier attachment
and tubing
Power Eclipse ON
Check tubing, cannula
and connections
Contact your dealer/provider
Repair or replace tubing or humidifier.
Place your Eclipse so there is adequate air flow.
Clean or replace air inlet filter. Place your Eclipse
so there is adequate air flow.
Switch to Continuous Flow Mode until respiratory
rate resumes to an acceptable rate.
Contact your dealer/provider
Repair or replace humidifier or tubing
Clean or replace air inlet filter
Contact your dealer/provider
Attach 7 foot (2.1m) tubing/cannula
Remove humidifier
Contact your dealer/provider
Switch to Pulse Mode
Place nasal cannula in nostrils
and breath through your nose
Contact your dealer/provider
to replace Power Cartridge (Battery)