Guidance and manufacturer's declaration – ELECTROMAGNETIC EMISSIONS –
Voltage
fluctuations/flicker
emissions
IEC 61000-3-3
Guidance and manufacturer's declaration – ELECTROMAGNETIC IMMUNITY –
The device is intended for use in the electromagnetic environment specified
below. The customer or the user of the device should assure that it is used in
such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
IACER Srl
FOR ALL EQUIPMENT AND SYSTEMS
Not
applicable
FOR ALL EQUIPMENT AND SYSTEMS
Test level
IEC 60601
±8kVa contact
±2kV; ±4kV;
±8kV; 15kV in
air
±2kV for
power supplies
lines
±1kV
input/output
±0.5kV, ±1kV
Line(s) to line
±0.5kV, ±1kV,
±2kV Line(s) to
earth
28
supplies buildings used for domestic
purposes.
Electromagnetic
Compliance
environment -
level
Floors should be
wood, concrete or
±8kVa
ceramic tile. If
contact
floors are covered
±2kV; ±4kV;
with synthetic
±8kV; 15kV
material, the
in air
relative humidity
should be at least
30%.
Mains power
quality should be
Not
that of a typical
applicable
commercial or
hospital
environment.
Mains power
quality should be
Not
that of a typical
applicable
commercial or
hospital
environment.
MNPG304