Guidance and manufacturer's declaration – ELECTROMAGNETIC IMMUNITY –
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
Note: UT is the A.C. mains voltage prior to application of the test level.
IACER Srl
FOR ALL EQUIPMENT AND SYSTEMS
0% UT
0,5 cycle
a 0°, 45°, 90°,
135°, 180°,
225°, 270° and
315°
0% UT
1 cycle
70% UT
25/30 cycle
Single phase
0% UT
At 0°
250/300 cycle
30 A/m
29
Main power
quality should be
that of a typical
commercial or
hospital
environment. If
the user of the
device requires
Not
continued
applicable
operation during
power mains
interruptions, it is
recommended
that the device be
powered from an
uninterruptible
power supply or a
battery.
Power frequency
magnetic fields
should be at levels
characteristic of a
30 A/m
typical location in
a typical
commercial or
hospital
environment
MNPG304