pH/Oxi 340i
Saving data
manually
Saving data
automatically
Int 1
Transmitting data
manually
Transmitting data
automatically
Int 2
Saving data
Press the <STO> key in the measuring mode (display No. with the number of
the next free memory location). Then, press <RUN/ENTER> and enter the ID
number with < > < >. Terminate the save with <RUN/ENTER>. The instru-
ment changes to the measuring mode.
The pH/Oxi 340i portable measuring instrument has an internal data memo-
ry. It can store up to 500 datasets. The save interval (Int 1) determines the
chronological interval between automatic save processes.
To set up the save interval, press <STO> while pressing the <RUN/ENTER>
key (display Int 1) and set the interval with < > < >key . Then, press <RUN/
ENTER> and enter the ID number with < > < >. Change to the last active
measuring mode with <RUN/ENTER>. The automatic save is switched on
(display Auto Store).
Outputting the data memory
You can output the data memory with the <RCL> key. By repeatedly press-
ing the <RCL> key, you reach the following functions:
StO dISP
Output stored data on the display
StO SEr
Output stored data on the serial interface
CAL dISP
Output calibration data on the display
CAL SEr
Output calibration data on the serial interface
Start the output with <RUN/ENTER>.
Transmitting data
Press the <RUN/ENTER> key in the measuring mode.
This manually triggers a data transmission of the current measured value to
the serial interface at any time - independently of the selected intervals.
The interval to the data transmission (Int 2) determines the chronological in-
terval between automatic data transmissions. After the selected interval ex-
pires, the current data record is transmitted to the interface. To set up the
transmission interval, press <RCL> while holding down the <RUN/ENTER>
key (display Int 2). Then set the interval with < > < >.
Saving data
11